How Flavanols Are Measured In A Lab

Cocoa flavanols are measured by HPLC: a weighed sample is extracted with an acidified solvent, the extract is pushed through a chromatography column that separates the flavanols by how many units are stuck together, and a fluorescence detector counts each fraction as it comes off. The figure a lab reports is not one reading. It is the sum of seven separate peaks, and the total moves depending on which official method the lab ran.

That last point is where the category gets sloppy. Brochures quote a flavanol figure as though it were a fixed property of the powder, like weight. It is a property of the powder and the method, and two accredited labs running two different official methods on one sample will hand you two different totals.

How flavanols are measured by HPLC, step by step

Flavanols are not volatile and cannot be counted by eye, so they are separated in solution and detected optically. Four stages.

  1. Extraction. The sample is defatted and the flavanols pulled into water, methanol or acetonitrile with a little acetic acid. Cocoa is roughly half fat, and skipping the defatting step wrecks the chromatography.
  2. Separation. The extract goes onto a column. The current AOAC method uses hydrophilic interaction chromatography, HILIC, on a diol column, which sorts the molecules by size. Run time is about 16 minutes.
  3. Detection. A fluorescence detector, not a UV detector. Flavanols fluoresce, most of the other things in a cocoa extract do not, so fluorescence buys specificity that a UV trace cannot.
  4. Calibration. Peak areas mean nothing until they are set against a standard of known concentration. Since 2020 that standard has been a cocoa flavanol reference material rather than a lab-made mixture.

DP1 to DP7, and why a total is a sum

Flavanols in cacao exist as single units and as chains. The jargon for chain length is degree of polymerisation, or DP. DP1 is the monomer, mostly epicatechin. DP2 to DP7 are procyanidins of two to seven units.

A "total cocoa flavanol" figure is those fractions added up. Which fractions you add up is a decision, not a measurement. Bussy, Ottaviani and Kwik-Uribe set this out in Food & Function in 2021: the two older methods summed DP1 to DP10, while AOAC 2020.05 sums DP1 to DP7. Same sample, different arithmetic, different answer.

Why the same sample gives three different numbers

The 2021 paper re-analysed stored samples across method generations and built regression models to translate between them. It is a methods comparison, not a trial, and the sample sets were small and clustered: 15 samples for the oldest comparison, 45 for the newest and 26 for the middle pair, with a separate 20-sample set used for cross-validation.

Method Fractions summed What it was built on
Pre-AOAC (Adamson) DP1 to DP10 A calibration mixture that is no longer available
AOAC 2012.24 (Robbins) DP1 to DP10 Normal-phase HPLC with fluorescence
AOAC 2020.05 (Bussy) DP1 to DP7 HILIC diol column, fluorescence, NIST RM 8403

The authors report that the gap between methods comes almost entirely from how the procyanidins, DP2 and above, are quantified rather than from the monomer. Two limitations are worth naming. The oldest calibrant no longer exists, so that comparison could not be cross-validated at all. And the authors are employed by Mars, which declares the conflict in the paper.

The reference material that keeps labs honest

In October 2020 NIST published the production report for RM 8403, a cocoa flavanol extract packed as 2 g sticks in aluminised mylar. Rimmer and colleagues assigned values fraction by fraction: 114.94 mg/g for DP1, 82.56 for DP2, down to 36.70 for DP7, each with an expanded uncertainty of roughly 1 to 2 mg/g.

Those values were not set by one lab. Nine laboratories ran the material in triplicate and the results were combined with a weighted-median model. Within-lab precision came out around 2 to 3% relative, between-lab around 2 to 4.5%. NIST is explicit that the between-lab figure is a lower bound on real reproducibility: nine labs following one protocol closely is a friendlier test than the open market.

Why a "total polyphenols" number is not a flavanol number

Plenty of cacao marketing quotes polyphenols in gallic acid equivalents. That figure almost always comes from the Folin-Ciocalteu assay, a colour-change reaction run in a plate reader. It is not chromatography and it identifies nothing.

Sánchez-Rangel and colleagues, writing in Analytical Methods in 2013, showed that reducing substances in a plant extract, ascorbic acid above all, react with the reagent and inflate the result. They proposed measuring that activity separately and subtracting it. Folin-Ciocalteu gives a bulk reducing-capacity index: useful for comparing batches of the same thing, useless as a flavanol figure.

What's actually in a Chaski tube

Chaski is a 600 g tube of 77% dark buttons. The settled part of the ingredient list is single-origin Peruvian and Ecuadorian cacao, coconut sugar, Lion's Mane as a fruiting body extract, and Cordyceps sinensis as mycelium. No emulsifiers, no seed oils, no palm oil, no added cane sugar. The cacao is low-temperature ground, and every batch is tested for lead, cadmium and pesticides.

Chaski has no measured cocoa flavanol figure. No batch has been through AOAC 2020.05 or any other flavanol method, so there is no number to publish and we do not publish one. Chaski makes no health claim of any kind. If you want to see the testing that does exist, the batch testing results for Chaski page lists what each batch is screened for. The full ingredient and format detail sits on the Chaski Cacao product page.

Two related pages go further into where flavanol figures move during manufacture: what alkalisation does to measured flavanol levels and how roasting temperature changes the numbers. The rest of the library is indexed in the Chaski guides hub.

Frequently Asked Questions

What does HPLC actually stand for and do?

High-performance liquid chromatography. A pump pushes a liquid sample through a packed column at high pressure. Different molecules stick to the packing material to different degrees, so they leave the far end at different times, and a detector records each one as a peak. For cocoa flavanols the current AOAC method separates on a diol column in hydrophilic interaction mode and detects with fluorescence, which responds to flavanols far more than to the rest of a cocoa extract. One run takes about 16 minutes. The output is a chromatogram, and the analyst converts peak areas into concentrations against a calibration standard.

Why do two labs report different flavanol levels for the same chocolate?

Usually because they ran different methods. Bussy and colleagues showed in 2021 that the older methods sum fractions DP1 to DP10 while AOAC 2020.05 sums DP1 to DP7, and that most of the divergence sits in how the longer procyanidin chains are quantified. Calibration differs too: the oldest approach used a lab-made mixture that is no longer available, while the 2020 method is anchored to NIST RM 8403. Sampling adds more spread, since a single bean lot is not uniform. So before comparing two figures, check the method name and the fractions included.

Is a polyphenol figure on a label the same as a flavanol figure?

No. Most polyphenol figures come from the Folin-Ciocalteu assay, a colorimetric reaction reported in gallic acid equivalents. It measures bulk reducing capacity, not specific compounds, and Sánchez-Rangel and colleagues documented in 2013 that other reducing substances in the extract, particularly ascorbic acid, push the number up. Flavanols are a subset of polyphenols, so a polyphenol total will always be the larger figure and cannot be read as a flavanol content. If a label quotes polyphenols, treat it as an index for comparing like with like, not as a flavanol measurement.

Does Chaski publish a flavanol number?

No. Chaski has no measured cocoa flavanol figure, because no batch has been analysed by AOAC 2020.05 or any earlier flavanol method. That is why you will not find a milligram figure for flavanols on the tube, on the product page or in these guides. The testing Chaski does run per batch is for lead, cadmium and pesticides.

Sources

  1. Bussy U, Ottaviani JI, Kwik-Uribe C, Food & Function, 2021
  2. Rimmer CA, Lippa KA, Yen J, Bussy U, Hewitt G, Anderson NP, Kwik-Uribe C, NIST Special Publication 260-207, 2020
  3. Sánchez-Rangel JC, Benavides J, Heredia JB, Cisneros-Zevallos L, Jacobo-Velázquez DA, Analytical Methods, 2013